Coulter counter

A Coulter counter is an apparatus for counting and sizing particles suspended in electrolytes. It is used for cells, bacteria, prokaryotic cells and virus particles.

Contents

A typical Coulter counter has one or more microchannels that separate two chambers containing electrolyte solutions. As fluid containing particles or cells is drawn through each microchannel, each particle causes a brief change to the electrical resistance of the liquid. The counter detects these changes in electrical resistance.

Coulter principle

The Coulter principle states that particles pulled through an orifice, concurrent with an electric current, produce a change in impedance that is proportional to the volume of the particle traversing the orifice. This pulse in impedance originates from the displacement of electrolyte caused by the particle. The Coulter principle was named for its inventor, Wallace H. Coulter. The principle has found commercial success in the medical industry, particularly in hematology, where it can be applied to count and size the various cells that make up whole blood.

Cells, being poorly conductive particles, alter the effective cross-section of the conductive microchannel. If these particles are less conductive than the surrounding liquid medium, the electrical resistance across the channel increases, causing the electric current passing across the channel to briefly decrease. By monitoring such pulses in electric current, the number of particles for a given volume of fluid can be counted. The size of the electric current change is related to the size of the particle, enabling a particle size distribution to be measured, which can be correlated to mobility, surface charge, and concentration of the particles.

The Coulter Counter is a vital constituent of today's hospital laboratory. Its primary function being the quick and accurate analysis of complete blood counts (often referred to as CBC). The CBC is used to determine the number or proportion of white and red blood cells in the body. Previously, this procedure involved preparing a blood cell stain and manually counting each type of cell under a microscope, a process that typically took a half-hour.

Coulter Counters have a wide variety of applications including paint, ceramics, glass, molten metals and food manufacture. They are also routinely employed for quality control.

A Coulter counter played an important role in the development of the first ever cell sorter, and was involved in the early days of the development of flow cytometry. Even today, some flow cytometers utilize the Coulter Principle to provide highly accurate information about cell size and count.

Many investigators have designed a variety of devices based on the Coulter Principle, and generated peer-reviewed publications featuring data generated by these devices. A few of these devices have also been commercialized. All implementations of the Coulter Principle feature trade offs between sensitivity, noise shielding, solvent compatibility, speed of measurement, sample volume, dynamic range, and reliability of device manufacture.

Development

Wallace H Coulter discovered the Coulter Principle in the late 1940s (though a patent was not awarded until October 20, 1953). Coulter was influenced by the atomic bombs dropped on Hiroshima and Nagasaki. These events motivated Coulter to simplify and improve blood cell analysis so that large populations could be screened rapidly, as would be necessary in the event of a nuclear war. Partial funding of the project came from a grant award from the Office of Naval Research.

"Coulter Principle" refers to the use of an electric field for counting and sizing dilute suspensions of particles in conducting liquids. Wallace H. Coulter was awarded US Patent #2,656,508, Means for Counting Particles Suspended in a Fluid. The Coulter Principle is most commonly employed in a Coulter counter, which is an analytical instrument designed for a specific task such as counting cells. However, there are numerous other ways to implement the Coulter Principle. Several of these have been attempted, some with commercial success, and some purely for academic research. To date, the most commercially successful application of the Coulter Principle is in hematology, where it is used to obtain information about patients’ blood cells.

The Coulter Principle relies on the fact that particles moving in an electric field cause measurable disturbances in that field. The magnitudes of these disturbances are proportional to the size of the particles in the field. Coulter identified several requirements necessary for practical application of this phenomenon. First, the particles should be suspended in a conducting liquid. Second, the electrical field should be physically constricted so that the movement of particles in the field causes detectable changes in the current. Finally, the particles should be dilute enough so that only one at a time passes through the physical constriction, preventing an artifact known as coincidence.

While the Coulter Principle can be implemented in a variety of designs, there are two that have become the most commercially relevant. These include an aperture format and a flow cell format. The Figure above shows several other geometries that Coulter patented.

Aperture format

The aperture format is used in most commercial Coulter counters. In this set up, an orifice of defined size is created in a jewel disk using special manufacturing processes. The resulting aperture is then embedded in the wall of a glass tube. The product of this process is commonly referred to as an aperture tube. While in use, the aperture tube is placed in a liquid so that the jewel disk is completely submerged and the tube can fill with liquid. Electrodes are positioned both inside and outside the aperture tube, which allows current to flow through the aperture. A pump is used to create a vacuum at the top of the tube, which draws the liquid through the aperture. Samples to be analyzed are then slowly added to the conducting liquid surrounding the aperture tube. At the start of the experiment, the electric field is turned on and the pump begins to draw the dilute suspension through the aperture. The resulting data are collected by recording the electrical pulses generated as the particles traverse the aperture.

While the basic physical set-up of the aperture format is consistent in every Coulter counter, the amount and quality of data varies greatly as a function of the signal processing circuitry implemented. For example, amplifiers with lower noise thresholds and greater dynamic range can increase the sensitivity of the system. Similarly, digital pulse height analyzers with variable bin widths provide much higher resolution data as opposed to analog analyzers with fixed bins. Further, combining a COULTER COUNTER® with a digital computer allows capture of many electrical pulse characteristics, while analog counters typically store a much more limited amount of information about each pulse.

Flow cell format

The flow cell format is most commonly implemented in hematology instruments, and sometimes flow cytometers. In this format, electrodes are embedded at either end of a flow channel and the electric field is applied through the channel. This format has several advantages as opposed to the aperture format. This arrangement allows for continuous sample analysis whereas the aperture format is single-batch format. Further, the use of a flow cell lends itself to addition of a sheath flow, which keeps particles centered in the middle of the flow channel. This allows measurements to be performed simultaneously, such as probing the object with a laser. The major disadvantages of the flow cell format are that it is much more expensive to manufacture and is typically fixed to one channel width, whereas the aperture format offers a wide variety of aperture sizes.

Coincidence

Anomalous electrical pulses can be generated if the concentration of sample is so high that multiple particles enter the aperture simultaneously. This situation is known as coincidence. This occurs because there is no way to ensure that a single large pulse is the result of a single large particle or multiple small particles entering the aperture at once. To prevent this situation, samples must be fairly dilute.

Particle path

The shape of the generated electrical pulse varies with the particle path through the aperture. Shoulders and other artifacts can occur because the electric field density varies across the diameter of the aperture. This variance is a result of both the physical constriction of the electric field and also the fact that the liquid velocity varies as a function of radial location in the aperture. In the flow cell format, this effect is minimized since sheath flow ensures each particle travels an almost identical path through the flow cell. In the aperture format, signal processing algorithms can be used to correct for artifacts resulting from particle path.

Conductive particles

Conductive particles are a common concern for individuals considering the Coulter Principle. While this topic raises interesting scientific questions, practically, it rarely affects the results of an experiment. This is because the conductivity difference between most conductive materials and ions in liquid (referred to as the discharge potential) is so great that most conductive materials act as insulators in a Coulter counter. The voltage necessary to break down this potential barrier is referred to as the breakdown voltage. For those highly conductive materials that present a problem, the voltage used during a Coulter experiment should be reduced below the breakdown potential (which can be determined empirically).

Porous particles

The Coulter Principle measures the volume of an object, since the disturbance in the electric field is proportional to the volume of electrolyte displaced from the aperture. This leads to some confusion amongst those who are used to optical measurements from microscopes or other systems that only view two dimensions and also show the boundaries of an object. The Coulter Principle, on the other hand measures three dimensions and the volume displaced by an object. It is most useful to think of sponges; even though a wet sponge may appear very large, it will displace significantly less liquid than a solid brick of the same dimensions.

Direct current vs alternating current

Direct current is used in the Coulter counters found in most research and cell laboratories. Direct current measurements are useful for an array of particles and allow for simplified data acquisition and processing. Alternating current measurements are sometimes used in clinical hematology instruments, due to the special nature of cell membranes. At low frequencies (below 500 kHz), alternating and direct current measurements behave essentially the same way. At intermediate frequencies (500 kHz - 6 MHz), the plasma membrane of cells can become polarized, leading to a decreased capacitance of the measurement systems. However, at high frequencies (6-20 MHz), the cell membrane loses its polarization, and the electrical pulses provide information about the cell cytoplasm.

Hematology

The most successful and important application of the Coulter Principle is in the characterization of human blood cells. The technique has been used to diagnose a variety of diseases, and is the standard method for obtaining red blood cell counts (RBCs) and white blood cell counts (WBCs) as well as several other common parameters. When combined with other technologies such as fluorescence tagging and light scattering, the Coulter Principle can help produce a detailed profile of patients’ blood cells.

Cell count and size

In addition to clinical counting of blood cells (cell diameters of ~6-10 micrometres, typically), the Coulter principle has established itself as the most reliable laboratory method for counting a wide variety of cells, ranging from bacteria (< 1 micrometre in size), fat cells (~400 micrometre), plant cell aggregates (>~1200 micrometre), and stem cell embryoid bodies (~900 micrometre). The technique has become so standardized that ASTM International has published a procedure on the topic: ASTMF2149-01(2007) Standard Test Method for Automated Analyses of Cells-the Electrical Sensing Zone Method of Enumerating and Sizing Single Cell Suspensions.

Particle characterization

The Coulter Principle has proved useful for applications well beyond cellular studies. The fact that it individually measures particles, is independent of any optical properties, is extremely sensitive, and is very reproducible has appeal to a wide variety of fields. Consequently, the Coulter Principle has been adapted to the nanoscale to produce a novel nanoparticle characterization technique called Tunable Resistive Pulse Sensing, or TRPS. TRPS enables high-fidelity analysis of a diverse set of nanoparticles, including (but not limited to): functionalized drug delivery nanoparticles, Virus-like particles (VLPs), liposomes, exosomes, polymeric nanoparticles, microbubbles.

To illustrate its versatility, below are listed some of the ASTM Methods that have been written for the Coulter Principle (Sometimes referred to as "Electrical Sensing Zone").

-Test Method C690-86 (1997) Standard Test Method for Particle Size Distribution of Alumina or Quartz by Electric Sensing Zone Technique

-Test Method E1772-95 Standard Test Method for Particle Size Distribution of Chromatography Media by Electric Sensing Zone Technique

-Test Method F751-83 (1997) Standard Test Method for Measuring Particle Size of Wide-Size Range Dry Toners

-Test Method F577-83 (1997) Standard Test Method for Particle Size Measurement of Dry Toners

-Test Method D4438-85 (1997) Standard Test Method for Particle Size Distribution of Catalytic Material by Electronic Counting

-Practice D3451-92 Standard Practices for Testing Polymeric

-Test Method F662-86 (1992)e1 Standard Test Method for Measurement of Particle Count and Size Distribution in Batch Samples for Filter Evaluation Using an Electrical Resistance Particle Counter

-Specification C757-90 (1996) Standard Specification for Nuclear-Grade Plutonium Dioxide Powder, Sinterable

Additionally, one International Organization for Standardization (ISO) guideline document has been produced on the Coulter Principle

-International Standard ISO 13319 Determination of particle size distributions —Electrical sensing zone method Reference number ISO 13319:2000(E)First edition 2000-04-01'