X-ray photoelectron spectroscopy (XPS) is a surface-sensitive quantitative spectroscopic technique that measures the elemental composition at the parts per thousand range, empirical formula, chemical state and electronic state of the elements that exist within a material. XPS spectra are obtained by irradiating a material with a beam of X-rays while simultaneously measuring the kinetic energy and number of electrons that escape from the top 0 to 10 nm of the material being analyzed. XPS requires high vacuum (P ~ 10−8 millibar) or ultra-high vacuum (UHV; P < 10−9 millibar) conditions, although a current area of development is ambient-pressure XPS, in which samples are analyzed at pressures of a few tens of millibar.
Contents
- Measurements
- History
- Basic physics
- Surface sensitivity
- Components of a commercial system
- Uses and capabilities
- Capabilities of advanced systems
- Chemical states and chemical shift
- Quantitative accuracy and precision
- Analysis time
- Detection limits
- Measured area
- Sample size limits
- Degradation during analysis
- General summary of use
- Materials routinely analyzed
- Charge compensation techniques
- Sample preparation
- Peak identification
- Charge referencing insulators
- Peak fitting
- FWHMs
- Chemical shifts
- Peak shapes
- Instrument design factors
- Experiment settings
- Sample factors
- Hemispherical electron energy analyzer
- Cylindrical mirror analyser
- Synchrotron based XPS
- Older style electron detector
- Quantum mechanical treatment
- Theory of core level photoemission of electrons
- References
XPS can be used to analyze the surface chemistry of a material in its as-received state, or after some treatment, for example: fracturing, cutting or scraping in air or UHV to expose the bulk chemistry, ion beam etching to clean off some or all of the surface contamination (with mild ion etching) or to intentionally expose deeper layers of the sample (with more extensive ion etching) in depth-profiling XPS, exposure to heat to study the changes due to heating, exposure to reactive gases or solutions, exposure to ion beam implant, exposure to ultraviolet light.
Measurements
XPS is used to measure:
XPS can be performed using a commercially built XPS system, a privately built XPS system, or a synchrotron-based light source combined with a custom-designed electron energy analyzer. Commercial XPS instruments in the year 2005 used either a focused 20- to 500-micrometer-diameter beam of monochromatic Al Kα X-rays, or a broad 10- to 30-mm-diameter beam of non-monochromatic (polychromatic) Al Kα X-rays or Mg Kα X-rays. A few specially designed XPS instruments can analyze volatile liquids or gases, or materials at pressures of roughly 1 torr (1.00 torr = 1.33 millibar), but there are relatively few of these types of XPS systems. The ability to heat or cool the sample during or prior to analysis is relatively common.
Because the energy of an X-ray with particular wavelength is known (for Al Kα X-rays, Ephoton = 1486.7 eV), and because the emitted electrons' kinetic energies are measured, the electron binding energy of each of the emitted electrons can be determined by using an equation that is based on the work of Ernest Rutherford (1914):
where Ebinding is the binding energy (BE) of the electron, Ephoton is the energy of the X-ray photons being used, Ekinetic is the kinetic energy of the electron as measured by the instrument and
History
In 1887, Heinrich Rudolf Hertz discovered but could not explain the photoelectric effect, which was later explained in 1905 by Albert Einstein (Nobel Prize in Physics 1921). Two years after Einstein's publication, in 1907, P.D. Innes experimented with a Röntgen tube, Helmholtz coils, a magnetic field hemisphere (an electron kinetic energy analyzer), and photographic plates, to record broad bands of emitted electrons as a function of velocity, in effect recording the first XPS spectrum. Other researchers, including Henry Moseley, Rawlinson and Robinson, independently performed various experiments to sort out the details in the broad bands.
After WWII, Kai Siegbahn and his research group in Uppsala (Sweden) developed several significant improvements in the equipment, and in 1954 recorded the first high-energy-resolution XPS spectrum of cleaved sodium chloride (NaCl), revealing the potential of XPS. A few years later in 1967, Siegbahn published a comprehensive study of XPS, bringing instant recognition of the utility of XPS, which he referred to as ESCA (Electron Spectroscopy for Chemical Analysis). In cooperation with Siegbahn, a small group of engineers (Mike Kelly, Charles Bryson, Lavier Faye, Robert Chaney) at Hewlett-Packard in the USA, produced the first commercial monochromatic XPS instrument in 1969. Siegbahn received the Nobel Prize for Physics in 1981, to acknowledge his extensive efforts to develop XPS into a useful analytical tool.
In parallel with Siegbahn's work, David Turner at Imperial College (and later at Oxford University) in the UK developed ultraviolet photoelectron spectroscopy (UPS) on molecular species using helium lamps.
Basic physics
A typical XPS spectrum is a plot of the number of electrons detected (sometimes per unit time) (Y-axis, ordinate) versus the binding energy of the electrons detected (X-axis, abscissa). Each element produces a characteristic set of XPS peaks at characteristic binding energy values that directly identify each element that exists in or on the surface of the material being analyzed. These characteristic spectral peaks correspond to the electron configuration of the electrons within the atoms, e.g., 1s, 2s, 2p, 3s, etc. The number of detected electrons in each of the characteristic peaks is directly related to the amount of element within the XPS sampling volume. To generate atomic percentage values, each raw XPS signal must be corrected by dividing its signal intensity (number of electrons detected) by a "relative sensitivity factor" (RSF), and normalized over all of the elements detected. Since hydrogen is not detected, these atomic percentages exclude hydrogen.
To count the number of electrons during the acquisition of a spectrum with a minimum of error, XPS detectors must be operated under ultra-high vacuum (UHV) conditions because electron counting detectors in XPS instruments are typically one meter away from the material irradiated with X-rays. This long path length for detection requires such low pressures.
Surface sensitivity
XPS detects only those electrons that have actually escaped from the sample into the vacuum of the instrument, and reach the detector. In order to escape from the sample into vacuum, a photoelectron must travel through the sample. Photo-emitted electrons can undergo inelastic collisions, recombination, excitation of the sample, recapture or trapping in various excited states within the material, all of which can reduce the number of escaping photoelectrons. These effects appear as an exponential attenuation function as the depth increases, making the signals detected from analytes at the surface much stronger than the signals detected from analytes deeper below the sample surface. Thus, the signal measured by XPS is an exponentially surface-weighted signal, and this fact can be used to estimate analyte depths in layered materials.
Components of a commercial system
The main components of a commercially made XPS system include a source of X-rays, an ultra-high vacuum (UHV) stainless steel chamber with UHV pumps, an electron collection lens, an electron energy analyzer, Mu-metal magnetic field shielding, an electron detector system, a moderate vacuum sample introduction chamber, sample mounts, a sample stage, and a set of stage manipulators.
Monochromatic aluminium K-alpha X-rays are normally produced by diffracting and focusing a beam of non-monochromatic X-rays off of a thin disc of natural, crystalline quartz with a <1010> orientation. The resulting wavelength is 8.3386 angstroms (0.83386 nm) which corresponds to a photon energy of 1486.7 eV. Aluminum K-alpha X-rays have an intrinsic FWHM of 0.43 eV, centered on 1486.7 eV (E/ΔE = 3457). For a well optimized monochromator, the energy width of the monochromated aluminum K-alpha X-rays is 0.16 eV, but energy broadening in common electron energy analyzers (spectrometers) produces an ultimate energy resolution on the order of FWHM=0.25 eV which, in effect, is the ultimate energy resolution of most commercial systems. When working under practical, everyday conditions, high-energy-resolution settings will produce peak widths (FWHM) between 0.4–0.6 eV for various pure elements and some compounds. For example, in a spectrum obtained in 1 minute at a pass energy of 20 eV using monochromated aluminum K-alpha X-rays, the Ag 3d5/2 peak for a clean silver film or foil will typically have a FWHM of 0.45 eV.
Non-monochromatic magnesium X-rays have a wavelength of 9.89 angstroms (0.989 nm) which corresponds to a photon energy of 1253 eV. The energy width of the non-monochromated X-ray is roughly 0.70 eV, which, in effect is the ultimate energy resolution of a system using non-monochromatic X-rays. Non-monochromatic X-ray sources do not use any crystals to diffract the X-rays which allows all primary X-rays lines and the full range of high-energy Bremsstrahlung X-rays (1–12 keV) to reach the surface. The ultimate energy resolution (FWHM) when using a non-monochromatic Mg K-alpha source is 0.9–1.0 eV, which includes some contribution from spectrometer-induced broadening.
Uses and capabilities
XPS is routinely used to determine:
Capabilities of advanced systems
Chemical states and chemical shift
The ability to produce chemical state information (as distinguished from merely elemental information) from the topmost few nm of any surface makes XPS a unique and valuable tool for understanding the chemistry of any surface, either as received, or after physical or chemical treatment(s). In this context, "chemical state" refers to the local bonding environment of a species in question. The local bonding environment of a species in question is affected by its formal oxidation state, the identity of its nearest-neighbor atom, its bonding hybridization to that nearest-neighbor atom, and in some cases even the bonding hybridization between the atom in question and the next-nearest-neighbor atom. Thus, while the nominal binding energy of the C 1s electron is 284.6 eV (some also use 285.0 eV as the nominal value for the binding energy of carbon), subtle but reproducible shifts in the actual binding energy, the so-called chemical shift, provide the chemical state information referred to here.
Chemical-state analysis is widely used for the element carbon. Chemical-state analysis of the surface of carbon-containing polymers readily reveals the presence or absence of the chemical states of carbon shown in bold, in approximate order of increasing binding energy, as: carbide (-C2−), silicone (-Si-CH3), methylene/methyl/hydrocarbon (-CH2-CH2-, CH3-CH2-, and -CH=CH-), amine (-CH2-NH2), alcohol (-C-OH), ketone (-C=O), organic ester (-COOR), carbonate (-CO32−), monofluoro-hydrocarbon (-CFH-CH2-), difluoro-hydrocarbon (-CF2-CH2-), and trifluorocarbon (-CH2-CF3), to name but a few examples.
Chemical state analysis of the surface of a silicon wafer readily reveals chemical shifts due to the presence or absence of the chemical states of silicon in its different formal oxidation states, such as: n-doped silicon and p-doped silicon (metallic silicon in figure above), silicon suboxide (Si2O), silicon monoxide (SiO), Si2O3, and silicon dioxide (SiO2). An example of this is seen in the figure above: High-resolution spectrum of an oxidized silicon wafer in the energy range of the Si 2p signal.
Quantitative accuracy and precision
Analysis time
Detection limits
Measured area
Sample size limits
Instruments accept small (mm range) and large samples (cm range), e.g. wafers. Limiting factor is the design of the sample holder, the sample transfer, and the size of the vacuum chamber. Large samples are laterally moved in x and y direction to analyse a larger area.
Degradation during analysis
General summary of use
Materials routinely analyzed
Inorganic compounds, metal alloys, semiconductors, polymers, pure elements, catalysts, glasses, ceramics, paints, papers, inks, woods, plant parts, make-up, teeth, bones, human implants, biomaterials, viscous oils, glues, ion modified materials
Charge compensation techniques
Sample preparation
Peak identification
The number of peaks produced by a single element varies from 1 to more than 20. Tables of binding energies (BEs) that identify the shell and spin-orbit of each peak produced by a given element are included with modern XPS instruments, and can be found in various handbooks [citations] and websites. Because these experimentally determined BEs are characteristic of specific elements, they can be directly used to identify experimentally measured peaks of a material with unknown elemental composition.
Before beginning the process of peak identification, the analyst must determine if the BEs of the unprocessed survey spectrum (0-1400 eV) have or have not been shifted due to a positive or negative surface charge. This is most often done by looking for two peaks that due to the presence of carbon and oxygen. {tbc}
Charge referencing insulators
Charge referencing is needed when a sample suffers either a positive (+) or negative (-) charge induced shift of experimental BEs. Charge referencing is needed to obtain meaningful BEs from both wide-scan, high sensitivity (low energy resolution) survey spectra (0-1100 eV), and also narrow-scan, chemical state (high energy resolution) spectra.
Charge induced shifting causes experimentally measured BEs of XPS peaks to appear at BEs that are greater or smaller than true BEs. Charge referencing is performed by adding or subtracting a "Charge Correction Factor" to each of the experimentally measured BEs. In general, the BE of the hydrocarbon peak of the C (1s) XPS signal is used to charge reference (charge correct) all BEs obtained from non-conductive (insulating) samples or conductors that have been deliberately insulated from the sample mount.
Charge induced shifting is normally due to: a modest excess of low voltage (-1 to -20 eV) electrons attached to the surface, or a modest shortage of electrons (+1 to +15 eV) within the top 1-12 nm of the sample caused by the loss of photo-emitted electrons. The degree of charging depends on various factors. If, by chance, the charging of the surface is excessively positive, then the spectrum might appear as a series of rolling hills, not sharp peaks as shown in the example spectrum.
The C (1s) BE of the hydrocarbon species (moieties) of the "Adventitious" carbon that appears on all, air-exposed, conductive and semi-conductive materials is normally found between 284.5 eV and 285.5 eV. For convenience, the C (1s) of hydrocarbon moieties is defined to appear between 284.6 eV and 285.0 eV. A value of 284.8 eV has become popular in recent years. However, some recent reports indicate that 284.9 eV or 285.0 eV represents hydrocarbons attached on metals, not the natural native oxide. The 284.8 eV BE is routinely used as the "Reference BE" for charge referencing insulators. When the C (1s) BE is used for charge referencing, then the charge correction factor is the difference between 284.8 eV and the experimentally measured C (1s) BE of the hydrocarbon moieties.
When using a monochromatic XPS system together with a low voltage electron flood gun for charge compensation the experimental BEs of the C (1s) hydrocarbon peak is often 4-5 eV smaller than the reference BE value (284.8 eV). In this case, all experimental BEs appear at lower BEs than expected and need to be increased by adding a value ranging from 4 to 5 eV. Non-monochromatic XPS systems are not usually equipped with a low voltage electron flood gun so the BEs will normally appear at higher BEs than expected. It is normal to subtract a charge correction factor from all BEs produced by a non-monochromatic XPS system.
Conductive materials and most native oxides of conductors should never need charge referencing. Conductive materials should never be charge referenced unless the topmost layer of the sample has a thick non-conductive film.
Peak-fitting
The process of peak-fitting high energy resolution XPS spectra is still a mixture of art, science, knowledge and experience. The peak-fit process is affected by instrument design, instrument components, experimental settings (aka analysis conditions) and sample variables. Most instrument parameters are constant while others depend on the choice of experimental settings.
Before starting any peak-fit effort, the analyst performing the peak-fit needs to know if the topmost 15 nm of the sample is expected to be a homogeneous material or is expected to be a mixture of materials. If the top 15 nm is a homogeneous material with only very minor amounts of adventitious carbon and adsorbed gases, then the analyst can use theoretical peak area ratios to enhance the peak-fitting process.
Variables that affect or define peak-fit results include:
FWHMs
When using high energy resolution experiment settings on an XPS equipped with a monochromatic Al K-alpha X-ray source, the FWHM of the major XPS peaks range from 0.3 eV to 1.7 eV. The following is a simple summary of FWHM from major XPS signals:However, it should be noted that the FWHM also depends on the detector, and can also increase due to the sample getting charged.
Chemical shifts
Chemical shift values depend on the degree of electron bond polarization between nearest neighbor atoms. A specific chemical shift is the difference in BE values of one specific chemical state versus the BE of one form of the pure element, or of a particular agreed-upon chemical state of that element. Component peaks derived from peak-fitting a raw chemical state spectrum can be assigned to the presence of different chemical states within the sampling volume of the sample.
Peak shapes
Depends on instrument parameters, experimental parameters and sample characteristicsInstrument design factors
FWHM and purity of X-rays used (monochromatic Al, non-monochromatic Mg, Synchrotron, Ag, Zr...)Design of electron analyzer (CMA, HSA, retarding field...)Experiment settings
Settings of the electron analyzer (e.g. pass energy, step size)Sample factors
Physical form of the sample (single crystal, polished, powder, corroded...)Number of physical defects within the analysis volume (from Argon ion etching, from laser cleaning...)Hemispherical electron energy analyzer
A hemispherical electron energy analyser is generally used for applications where a higher resolution is needed. An ideal hemispherical analyser consists of two concentric hemispherical electrodes (inner and outer hemispheres) held at proper voltages. It is possible to demonstrate that in such a system, (i) the electrons are linearly dispersed along the direction connecting the entrance and the exit slit, depending on their kinetic energy, while (ii) electrons with the same energy are first-order focused. When two potentials,
where
where
The latter equation can be used to determine the potentials to be applied to the hemispheres in order to select electrons with energy
In fact, only the electrons with energy
The instrumental energy resolution of the device depends both on the geometrical parameters of the analyzer and on the angular divergence of the incoming photoelectrons:
where
When acquiring spectra in sweep (or scanning) mode, the voltages of the two hemispheres
Cylindrical mirror analyser
Since the relevant information, in photoemission spectroscopy, is contained in the kinetic energy distribution of the photoelectrons, a specific device is needed to energy-filter the electrons emitted (or scattered) by the sample. Electrostatic monochromators are the most common choice. The older design, a CMA, represents a trade-off between the need for high count rates and high angular/energy resolution. The so-called cylindrical mirror analyser (CMA) is mostly used for checking the elemental composition of the surface. It consists of two co-axial cylinders placed in front of the sample, the inner one being held at a positive potential, while the outer cylinder is held at a negative potential. Only the electrons with the right energy can pass through this set-up and are detected at the end. The count rates are high but the resolution (both in energy and angle) is poor.
Synchrotron based XPS
A breakthrough has been actually brought about in the last decades by the development of large scale synchrotron radiation facilities. Here, bunches of relativistic electrons kept on a circular orbit inside a storage ring are accelerated through bending magnets or insertion devices like wigglers and undulators to produce a high brilliance and high flux photon beam. The main advantages of using synchrotron light are
The highest spectral brightness and narrowest beam energy dispersion is attained by undulators, which consist of periodic array of dipole magnets in which the electrons are forced to wiggle and thus to emit coherent light. Besides the high intensity, energy tunability is one of the most important advantages of synchrotron light compared to the light produced by conventional X-ray sources. In fact, a wide energy range (from the IR to the Hard X-ray region, depending on the energy of the electron bunch) is accessible by changing the undulator gaps between the arrays. Continuous energy spectra available from a synchrotron radiation source allows selection of photon energies yielding optimum photoionization cross-sections appropriate for probing a particular core level. The high photon flux, in addition, makes it possible to perform XPS experiments also from low density atomic species, such as molecular and atomic adsorbates.
Older style electron detector
Electrons can be detected using an electron multiplier, usually a channeltron. This device essentially consists of a glass tub with a resistive coating on the inside. A high voltage is applied between the front and the end. An electron which enters the channeltron is accelerated to the wall, where it removes more electrons, in such a way that an electron avalanche is created, until a measurable current pulse is obtained.
Quantum mechanical treatment
When a photoemission event takes place, the following energy conservation rule holds:
where
From the theoretical point of view, the photoemission process from a solid can be described with a semiclassical approach, where the electromagnetic field is still treated classically, while a quantum-mechanical description is used for matter. The one—particle Hamiltonian for an electron subjected to an electromagnetic field is given by:
where
Actually, neglecting the
In first-order perturbation approach, the one-electron Hamiltonian can be split into two terms, an unperturbed Hamiltonian
In the time-dependent perturbation theory for harmonic perturbations, the transition rate between the initial state
where
In a real photoemission experiment the ground state core electron BE cannot be directly probed, because the measured BE incorporates both initial state and final state effects, and the spectral linewidth is broadened owing to the finite core-hole lifetime (
Assuming an exponential decay probability for the core hole in the time domain (
From the theory of Fourier transforms,
The photoemission event leaves the atom in a highly excited core ionized state, from which it can decay radiatively (fluorescence) or non-radiatively (typically by Auger decay). Besides Lorentzian broadening, photoemission spectra are also affected by a Gaussian broadening, whose contribution can be expressed by
Three main factors enter the Gaussian broadening of the spectra: the experimental energy resolution, vibrational and inhomogeneous broadening. The first effect is caused by the non perfect monochromaticity of the photon beam -which results in a finite bandwidth- and by the limited resolving power of the analyser. The vibrational component is produced by the excitation of low energy vibrational modes both in the initial and in the final state. Finally, inhomogeneous broadening can originate from the presence of unresolved core level components in the spectrum.
Theory of core level photoemission of electrons
In a solid, also inelastic scattering events contribute to the photoemission process, generating electron-hole pairs which show up as an inelastic tail on the high BE side of the main photoemission peak. In some cases, we observe also energy loss features due to plasmon excitations. This can either a final state effect caused by core hole decay, which generates quantized electron wave excitations in the solid (intrinsic plasmons), or it can be due to excitations induced by photoelectrons travelling from the emitter to the surface (extrinsic plasmons). Due to the reduced coordination number of first-layer atoms, the plasma frequency of bulk and surface atoms are related by the following equation:
Vibrational effects
Temperature-dependent atomic lattice vibrations, or phonons, can broaden the core level components and attenuate the interference patterns in an XPD (X-Ray Photoelectron Diffraction) experiment. The simplest way to account for vibrational effects is by multiplying the scattered single-photoelectron wave function
where