A Faraday cup is a metal (conductive) cup designed to catch charged particles in vacuum. The resulting current can be measured and used to determine the number of ions or electrons hitting the cup. The Faraday cup is named after Michael Faraday who first theorized ions around 1830.

Contents

Principle of operation

When a beam or packet of ions hits the metal it gains a small net charge while the ions are neutralized. The metal can then be discharged to measure a small current equivalent to the number of impinging ions. Essentially the Faraday cup is part of a circuit where ions are the charge carriers in vacuum and the faraday cup is the interface to the solid metal where electrons act as the charge carriers (as in most circuits). By measuring the electric current (the number of electrons flowing through the circuit per second) in the metal part of the circuit the number of charges being carried by the ions in the vacuum part of the circuit can be determined. For a continuous beam of ions (each with a single charge)

N t = I e

where N is the number of ions observed in a time t (in seconds), I is the measured current (in amperes) and e is the elementary charge (about 1.60 × 10−19 C). Thus, a measured current of one nanoamp (10−9 A) corresponds to about 6 billion ions striking the faraday cup each second.

Similarly, a Faraday cup can act as a collector for electrons in a vacuum (for instance from an electron beam). In this case electrons simply hit the metal plate/cup and a current is produced. Faraday cups are not as sensitive as electron multiplier detectors, but are highly regarded for accuracy because of the direct relation between the measured current and number of ions.This device is considered a universal charge detector because of its independence from the energy, mass, chemistry, etc. of the analyte.

Faraday cup in plasma diagnostics

Faraday cup utilizes a physical principle according which the electrical charges delivered to the inner surface of a hollow conductor are redistributed around its outer surface due to mutual self-repelling of charges of the same sign – phenomenon discovered by Faraday.

The conventional Faraday cup is applied for measurements of ion (or electron) flows from plasma boundaries and comprises a metallic cylindrical receiver-cap – 1 (Fig. 1) closed with, and insulated from, a washer-type metallic electron-suppressor lid - 2 provided with the round axial through enter-hollow of an aperture with a surface area S F = π D F 2 / 4 . Both the receiver cup and the electron-suppressor lid are enveloped in, and insulated from, a grounded cylindrical shield - 3 having an axial round hole coinciding with the hole in the electron-suppressor lid - 2. The electron-suppressor lid is connected by 50 Ω RF cable with the source B e s of variable DC voltage U e s . The receiver-cup is connected by 50 Ω RF cable through the load resistor R F with a sweep generator producing saw-type pulses U g ( t ) . Electric capacity C F is formed of the capacity of the receiver-cup - 1 to the grounded shield - 3 and the capacity of the RF cable. The signal from R F enables an observer to acquire an I-V characteristic of the Faraday cup by oscilloscope. Proper operating conditions: h D F (due to possible potential sag) and h λ i , where λ i is the ion free path. Signal from R F is the Faraday cup I-V characteristic which can be observed and memorized by oscilloscope

i Σ ( U g ) = i i ( U g ) C F d U g d t . (1)

In Fig. 1: 1 – cup-receiver, metal (stainless steel). 2 – electron-suppressor lid, metal (stainless steel). 3 – grounded shield, metal (stainless steel). 4 - insulator (teflon, ceramic). C F - capacity of Faraday cup. R F - load resistor.

Thus we measure the sum i Σ of the electric currents through the load resistor R F : i i (Faraday cup current) plus the current i c ( U g ) = C F ( d U g / d t ) induced through the capacitor C F by the saw-type voltage U g of the sweep-generator: The current component i c ( U g ) can be measured at the absence of the ion flow and can be subtracted further from the total current i Σ ( U g ) measured with plasma to obtain the actual Faraday cup I-V characteristic i i ( U g ) for processing. All of the Faraday cup elements and their assembly that interact with plasma are fabricated usually of temperature-resistant materials (often these are stainless steel and teflon or ceramic for insulators). For processing of the Faraday cup I-V characteristic, we are going to assume that the Faraday cup is installed far enough away from an investigated plasma source where the flow of ions could be considered as the flow of particles with parallel velocities directed exactly along the Faraday cup axis. In this case, the elementary particle current d i i corresponding to the ion density differential d n ( v ) in the range of velocities between v and v + d v of ions flowing in through operating aperture S F of the electron-suppressor can be written in the form

d i i = e Z i S F v d n ( v ) , (2)

where

d n ( v ) = n f ( v ) d v , (3)

e is elementary electric charge, Z i is the ion charge state, and f ( v ) is the one-dimensional distribution function of ions over velocity v . Therefore the ion current at the ion-decelerating voltage U g of the Faraday cup can be calculated by integrating Eq. (2) after substituting in it Eq. (3)

i i ( U g ) = e Z i n i S F 2 e Z i U g / M i f ( v ) v d v , (4)

where the lower integration limit is defined from the obvious equation M i v i , s 2 / 2 = e Z i U g where v i , s is the velocity of the ion stopped by the decelerating potential U g , and M i is the ion mass. Thus the expression (4) represents the I-V characteristic of the Faraday cup. Differentiating Eq. (4) with respect to U g , one can obtain the relation

d i i ( U g ) d U g = e n i S F e Z i M i f ( 2 e Z i U g / M i ) , (5)

where the value n i S F ( e Z i / M i ) = C i is an invariable constant for each measurement. Therefore the average velocity v i of ions arriving into the Faraday cup and their average energy E i can be calculated (under the assumption that we operate with a single type of ion) by the expressions

v i = 1.389 × 10 6 Z i M A 0 i i ( U g ) d U g ( 0 i i U g d U g ) 1 [cm/s], (6)

E i = 0 i i ( U g ) U g d U g ( 0 i i U g d U g ) 1 [eV], (7)

where M A is the ion mass in atomic units. The ion concentration n i in the ion flow at the Faraday cup vicinity can be calculated by the formula

n i = i i ( 0 ) e Z i v i S F (8)

which follows from Eq. (4) at U g = 0 ,

0 f ( v ) v d v = v , (9)

and from the conventional condition for distribution function normalizing

0 f ( v ) d v = 1 . (10)

Fig. 2 illustrates the I-V characteristic i i ( V ) and its first derivative i i ( V ) of the Faraday cup with S F = 0.5 c m 2 installed at output of the Inductively coupled plasma source powered with RF 13.56 MHz and operating at 6 mTorr of H2. The value of the electron-suppressor voltage (accelerating the ions) was set experimentally at U e s = 170 V , near the point of suppression of the secondary electron emission from the inner surface of the Faraday cup.